First page Back Continue Last page Overview Graphics
Multi-slice dose issues: paediatric protocols
More effort in defining paediatric CT protocols
- for same mAs, dose to newborn is up to 2.5 that of adult
- image quality has to be considered
Siemens system modifies the tube current for paediatrics
- based upon
- patient weight for head exams
- patient age for bodies
GE uses ‘rainbow system’ for
colour coding different age groups
- acquisition parameters set for
each colour
- rainbow system extends
to drug doses etc