First page Back Continue Last page Overview Graphics
Cardiac Gating
Cardiac movement normally blurs CT images
Using a pitchx of ~0.25 gives overlapping spirals, each part of the heart imaged for ~ four rotations
Four data sets for each slice, each starting at a different cardiac phase
Using ECG, data sets can be combined so that all projection angles are covered for a similar cardiac phase