Slide 5 of 13
Notes:
- Image noise and uniformity were measured using body and head phantoms.
- We used water filled phantoms
- Our body phantom was made of 32-cm diameter cylindrical Perspex shell filled with water.
- And the Head phantom was made of 16-cm diameter bone equivalent cylindrical shell with Perspex ends.
Both phantoms have air traps designed to trap any air bubbles out side the scanned volume.